Draeger Tube Acrylonitrile 0.2/a 8103701
Original price was: $150.00.$135.00Current price is: $135.00.
The Draeger Tube Acrylonitrile 0.2/a 8103701 has a measurement range of 0.2 – 4 ppm 5 – 50 ppm. During measurements, color changes on the tube indicate the concentration level of the hazardous substance.
- Box of 10 tubes
- 2 year shelf life
Description
The Draeger Tube Acrylonitrile 0.2/a 8103701 is high precision. The Dräger sampling tubes and systems provide very reliable results, even with complex compounds and mixtures of substances.
Application Range
- Standard Measuring Range: 0.2 to 4 ppm / 5 to 50 ppm
- Number of Strokes n: 20 / 5
- Time for Measurement: approx. 4 min / approx. 1 min
- Standard Deviation: ± 15 to 20 %
- Color Change: yellow → red
Ambient Operating Conditions
- Temperature: 5 to 40 °C
- Absolute Humidity: 1 to 25 mg H2O / L
Reaction Principle
a. CH2=CH-CN + MnO4 → HCN
b1. HCN + HgCl2 → HCl
b2. HCl + methyl red → red reaction product
Cross Sensitivity
At 4 ppm acrylonitrile no effect from:
- 1000 ppm acetone
- 20 ppm benzene
- 1000 ppm, ethal acetate
The tube indicates acrylonitrile with lower sensitivity and determining the concetration is not possible in the presence of:
- 500 ppm ethanol
- 1000 ppm n-hexane
- 100 ppm toluene
In the presence of 400 ppm butadiene, the indication of 4 ppm acrylonitrile is largely suppressed.
Draeger Acrylonitrile 0.2/a Tubes – 8103701
- Simple to use
- Requires no calibration
- Inexpensive and portable
- Wide variety of gases and vapors
- 70 years of proven technology
- Most Trusted – Draeger-Tubes have been leading the way in gas detection for over 60 years.
- Widest Range – Over 200 different Draeger-Tubes are available for measuring over 500 gases and vapors.
- Hassle-Free Operation – No calibration, no lubrication, no special maintenance. Always ready to go.
Accurate and reliable results
Draeger-Tubes with the Accuro Pump deliver the most accurate results. 60+ years of manufacturing colorimetric tubes. The bellows pump design delivers consistent volume. Individual calibration of each batch of Draeger gas detection tubes Acrylonitrile included, accomplishes quality assurance. Every batch undergoes testing at regular intervals, throughout the two year shelf life, to guarantee accuracy over the entire life of the tubes.
Draeger Acrylonitrile Detector Tubes
Collect the hazardous substances in the air by using a suitable medium such as activated carbon, silica gel, or molecular sieves. Then analyze the sample in a laboratory with the aid of instrumental analysis.
Suitable for Complex Cases
Draeger sampling tubes and systems allow identification and measurement of different substances even under difficult conditions. Use Draeger Tubes to simultaneously and accurately measure very complex mixtures of substances or cases in which several similar substances are present. Dräger Tube Acrylonitrile 0.2/a has the Draeger SKU 8103701.
Dräger accuro
Fast measurement with one hand: The Dräger-Tube pump accuro allows you to use the established Dräger-Tubes to take measurements under extreme conditions. The easy to use Draeger Tube 8103701 Acrylonitrile 0.2/a has already undergone calibration. It forms a perfect unit with the Dräger-Tube pump including the Draeger Safety Acrylonitrile Tube.
- Dräger Accuro Pump info here >>
- Draeger Hard-Side Accuro Pump Kit info here >>
Acrylonitrile Draeger Tubes & CMS Handbook |
App Store Draeger Tubes App |
Play Store Draeger Tubes App |
What are Draeger Tubes?
Dräger-Tubes are glass vials that contain a chemical reagent that reacts to a specific chemical or family of chemicals. Draeger detector tubes are a cost-effective, fast, and reliable way to measure toxic gases anywhere.
Draeger Tube Acrylonitrile 0.2/a Gas Detection Tube 8103701
For more information about this product, contact us today +1(805) 617-0987.
Additional information
Weight | 1 lbs |
---|---|
Dimensions | 7 × 2 × 2 in |