Dräger-Tubes for Measuring Air Samples
Application Range
- Standard Measuring Range: 0.25 to 3 ppm
- Number of Strokes n: 10
- Time for Measurement: approx. 1 min
- Standard Deviation: ± 10 to 15 %
- Color Change: yellow → blue
Ambient Operating Conditions
- Temperature: 10 to 50 °C
- Absolute Humidity: < 20 mg H2O / L
Reaction Principle
- NH3 + pH indicator → blue reaction product
Cross Sensitivity
Other basic substances such as e.g. organic amines are indicated as well, but with different sensitivity
Dräger-Tubes for Measuring Liquid Samples
Application Range
Determination of Draeger ammonia tubes in water/waste water
- Dräger-Tube: Ammonia 0.25/a
- Measuring range: 1.5 to 10 mg/L
- Number of Strokes (n): 10
- Typical Stroke Time: 10 to 30 s
- Measurement Time: approx. 200 s
- Sample Volume: 200 mL
- Color Change: yellow → blue
- Temperature Range: 4 to 30 °C
- pH-Measurement: necessary
Information of Measurement
Using acetic acid or sodium hydroxide solution, the pH-value has to be adjusted to the value of 10.2 – 10.3.
System Parameters (valid for pH 1.3)
Measurement Range [mg/L] | Standard Deviation [%] | Temperature [°C] | Parameters B C |
1.5 to 10 | 30 | 4 to 7
8 to 12 13 to 17 18 to 24 25 to 30 |
3.427 2.926
2.578 1.895 1.397 1.409 0.815 0.918 0.989 0.774 |
Evaluation of Measurement
Calculate ammonia concentration:
Y[mg/L] = A • B • (X[ppm] + C)
Cross Sensitivity
Other basic substances are also indicated. Draeger Tube Ammonia 1.5 to 10 mg/L 8101711
Draeger Tube Ammonia 10 to 100 mg/L 8101711
Application Range
Determination of Draeger ammonia tubes in water/waste water
- Dräger-Tube: Ammonia 0.25/a
- Measuring range: 10 to 100 mg/L
- Number of Strokes (n): 1
- Typical Stroke Time: 10 to 30 s
- Measurement Time: approx. 20 s
- Sample Volume: 200 mL
- Color Change: yellow → blue
- Temperature Range: 4 to 30 °C
- pH-Measurement: necessary
Information of Measurement
Using acetic acid or sodium hydroxide solution, the pH-value has to be adjusted to the value of 10.2 – 10.3.
System Parameters (valid for pH 1.3)
Measurement Range [mg/L] | Standard Deviation [%] | Temperature [°C] | Parameters B C |
10 to 100 | 30 | 4 to 7
8 to 12 13 to 17 18 to 24 25 to 30 |
61.34 0.826
40.46 0.310 29.37 0.943 27.59 0.463 18.11 -0.123 |
Evaluation of Measurement
Calculate ammonia concentration:
Y[mg/L] = A • B • (X[ppm] + C)
Cross Sensitivity
Other basic substances are also indicated.
Draeger Ammonia 0.25/a Tubes – 8101711
- Simple to use
- Requires no calibration
- Inexpensive and portable
- Wide variety of gases and vapors
- 70 years of proven technology
- Most Trusted – Draeger-Tubes have been leading the way in gas detection for over 60 years.
- Widest Range – Over 200 different Draeger-Tubes are available for measuring over 500 gases and vapors.
- Hassle-Free Operation – No calibration, no lubrication, no special maintenance. Always ready to go.
Accurate and reliable results
Draeger-Tubes with the Accuro Pump deliver the most accurate results. 60+ years of manufacturing colorimetric tubes. The bellows pump design delivers consistent volume. Individual calibration of each batch of Draeger gas detection tubes Ammonia included, accomplishes quality assurance. Every batch undergoes testing at regular intervals, throughout the two year shelf life, to guarantee accuracy over the entire life of the tubes.
Draeger Ammonia Detector Tubes
Collect the hazardous substances in the air by using a suitable medium such as activated carbon, silica gel, or molecular sieves. Then analyze the sample in a laboratory with the aid of instrumental analysis.
Suitable for complex cases
Draeger sampling tubes and systems allow identification and measurement of different substances even under difficult conditions. Use Draeger Tubes to simultaneously and accurately measure very complex mixtures of substances or cases in which several similar substances are present. Dräger Tube Ammonia 0.25/a has the Draeger SKU 8101711.
Dräger accuro
Fast measurement with one hand: The Dräger-Tube pump accuro allows you to use the established Dräger-Tubes to take measurements under extreme conditions. The easy to use Draeger Tube 8101711 Ammonia 0.25/a has already undergone calibration. It forms a perfect unit with the Dräger-Tube pump including the Draeger Safety Ammonia Tube.
- Dräger Accuro Pump info here >>
- Draeger Hard-Side Accuro Pump Kit info here >>
Ammonia Draeger Tubes & CMS Handbook | App Store – Draeger Tubes App | Play Store – Draeger Tubes App |
What are Draeger Tubes?
Dräger-Tubes are glass vials that contain a chemical reagent that reacts to a specific chemical or family of chemicals. Draeger detector tubes are a cost-effective, fast, and reliable way to measure toxic gases anywhere.
Draeger Tube Ammonia 0.25/a Gas Detection Tube 8101711
For more information about this product, contact us today +1(805) 617-0987.
Additional information
Weight | 1 lbs |
---|---|
Dimensions | 7 × 2 × 2 in |