Draeger Tube Perchloroethylene 2/a 8101501
$85.50
The Draeger Tube Perchloroethylene 2/a 8101501 has a measurement range of 20 to 300 ppm / 2 to 40 ppm. During measurements, color changes on the tube indicate the concentration level of the hazardous substance.
- Box of 10 tubes
- 2 year shelf life
Description
The Draeger Tube Perchloroethylene 2/a 8101501 is high precision. The Dräger sampling tubes and systems provide very reliable results, even with complex compounds and mixtures of substances and used for multiple chemicals in air, liquid, oil and soil.
Dräger-Tubes for Measuring Air Samples
Application Range
- Standard Measuring Range: 20 to 300 ppm / 2 to 40 ppm
- Number of Strokes n: 1 / 5
- Time for Measurement: approx. 30 s / approx. 3 min
- Standard Deviation: ± 15 to 20 %
- Color Change: yellow → grey blue
Ambient Operating Conditions
- Temperature: 15 to 30 °C
- Absolute Humidity: < 25 mg H2O / L
Reaction Principle
- CCI2=CCI2 + MNO4- → CI2
- CI2 + diphenylbenzidine → grey blue reaction product
At higher concentrations there may be created a reddish area at the beginning of the indicating layer
Cross Sensitivity
Other chlorinated hydrocarbons, free halogens and hydrogen halides are also indicated. Petroleum hydrocarbon vapors result in a shortened reading once they exceed the following concentrations: 50 ppm with 5 strokes or 500 ppm with 1 stroke.
Dräger-Tubes for Measuring Multiple Phase Samples
Application Range
Determination of volatile chlorinated hydrocarbons in multiple phase
- Dräger-Tube: Perchloroethylene 2/a
- Measuring range: qualitative
- Number of Strokes (n): maximum 10
- Typical Stroke Time: 45 to 65 s
- Measurement Time: approx. 55 to 550 s
- Sample Volume: 200 mL
- Color Change: yellow white → grey blue
- Temperature Range: 10 to 25 °C
- pH-Measurement: not necessary
Information of Measurement
– Mix a multiple phase sample which consists of e.g. 250 g water, 10 g fixed phase and 10 g oil part (about 300 mL) is mixed with approx. 5 g activated coal. It must rest for 3 minutes and then be shaken for 1 min.
– 0.2 g hydrophobated peat is added and the it must be shaken for 1 minute.
– The liquid is filled into the gas wash bottle up to the 200 mL mark.
Evaluation of Measurement
The measurement evaluation is qualitative (yes or no)
Cross Sensitivity
Chlorobenzene, 1,1-dichloroethane, 1,2-dichloroethane, dichloro methane, perchloroethylene, trichloroethylene and trichloromethane are indicated. Carbon tetrachloride and 1,1,1-trichloroethane are not indicated. Draeger tube Chlorinated Hydrocarbons qualitative in multiple phase 8101501
Dräger-Tubes for Measuring Oil Samples
Application Range
Determination of volatile chlorinated hydrocarbons in oil sludges/oil emulsions
- Dräger-Tube: Perchloroethylene 2/a
- Measuring range: qualitative
- Number of Strokes (n): maximum 10
- Typical Stroke Time: 45 to 65 s
- Measurement Time: approx. 55 to 550 s
- Sample Volume: approx. 0.5 g
- Color Change: yellow white → grey blue
- Temperature Range: 10 to 25 °C
- pH-Measurement: not necessary
Information of Measurement
– Approx. 0.5 g oil sample has to be shaken intensively with 1 L de-ionized water for 2 minutes in a laboratory bottle.
– The solution must be filtered through an analysis funnel with a round filter (black ribbon) directly into the gas wash bottle up to the 200 mL mark.
Evaluation of Measurement
The measurement evaluation is qualitative (yes or no)
Cross Sensitivity
Chlorobenzene, 1,1-dichloroethane, 1,2-dichloroethane, dichloro -methane, perchloroethylene, trichloroethylene and trichloromethane are indicated. Carbon tetrachloride and 1,1,1-trichloroethane are not indicated. Draeger tubes Chlorinated Hydrocarbons qualitative in oil 8101501
Dräger-Tubes for Measuring Liquid Samples
Application Range
Determination of perchloroethylene in water/waste water
- Dräger-Tube: Perchloroethylene 2/a
- Measuring range: 0.1 to 1 mg/L / 0.5 to 2 mg/L
- Number of Strokes (n): 8 / 4
- Typical Stroke Time: 45 to 65 s
- Measurement Time: approx. 440 s / ca. 220 s
- Sample Volume: 200 mL
- Color Change: yellow white → grey blue
- Temperature Range: 8 to 37 °C
- pH-Measurement: not necessary
System Parameters
Measurement Range [mg/L] | Standard Deviation [%] | Temperature [°C] | Parameters B C |
0.1 to 1 number of strokes n=8 | 25
20 20 20 20 25 |
8 to 12
13 to 17 18 to 22 23 to 27 28 to 32 33 to 37 |
0.035 0
0.031 0 0.028 0 0.026 0 0.025 0 0.023 0 |
0.5 to 2 number of strokes n=4 | 25
20 20 20 25 30 |
8 to 12
13 to 17 18 to 22 23 to 27 28 to 32 33 to 37 |
0.075 0
0.071 0 0.065 0 0.057 0 0.056 0 0.047 0 |
Evaluation of Measurement
Calculate perchloroethylene concentration:
Y[mg/L] = A • B • (X[ppm] + C)
Cross Sensitivity
Dichloromethane and chloroform are indicated with lower sensitivity. Trichloroethylene is indicated with nearly the same sensitivity. Petroleum hydrocarbons, benzene, carbon tetrachloride, toluene, 1,1,1-trichloroethane and xylene are not indicated. Draeger Tube Perchloroethylene 0.1 to 2 mg/L 8101501
Draeger Tube Trichloroethylene 0.1 to 1 mg/L 8101501
Application Range
Determination of trichloroethylene in water/waste water
- Dräger-Tube: Perchloroethylene 2/a
- Measuring range: 0.1 to 1 mg/L
- Number of Strokes (n): 8
- Typical Stroke Time: 45 to 65 s
- Measurement Time: approx. 440 s
- Sample Volume: 200 mL
- Color Change: yellow white → grey blue
- Temperature Range: 5 to 33 °C
- pH-Measurement: not necessary
System Parameters
Measurement Range [mg/L] | Standard Deviation [%] | Temperature [°C] | Parameters B C |
0.1 to 1 | 30 | 5 to 10
11 to 15 16 to 22 23 to 28 29 to 33 |
0.033 0
0.030 0 0.024 0 0.020 0 0.018 0 |
Evaluation of Measurement
Calculate trichloroethylene concentration:
Y[mg/L] = A • B • (X[ppm] + C)
Cross Sensitivity
Dichloromethane, chlorobenzene, 1,1-dichloroethane, 1,2 dichloroethane and chloroform are indicated with lower sensitivity. Perchloroethylene is indicated with nearly the same sensitivity. Carbon tetrachloride and 1,1,1-trichloroethane are not indicated.
Draeger Perchloroethylene 2/a Tubes – 8101501
- Simple to use
- Requires no calibration
- Inexpensive and portable
- Wide variety of gases and vapors
- 70 years of proven technology
- Most Trusted – Draeger-Tubes have been leading the way in gas detection for over 60 years.
- Widest Range – Over 200 different Draeger-Tubes are available for measuring over 500 gases and vapors.
- Hassle-Free Operation – No calibration, no lubrication, no special maintenance. Always ready to go.
Accurate and reliable results
Draeger-Tubes with the Accuro Pump deliver the most accurate results. 60+ years of manufacturing colorimetric tubes. The bellows pump design delivers consistent volume. Individual calibration of each batch of Draeger gas detection tubes Perchloroethylene included, accomplishes quality assurance. Every batch undergoes testing at regular intervals, throughout the two year shelf life, to guarantee accuracy over the entire life of the tubes.
Draeger Perchloroethylene Detector Tubes
Collect the hazardous substances in the air by using a suitable medium such as activated carbon, silica gel, or molecular sieves. Then analyze the sample in a laboratory with the aid of instrumental analysis.
Suitable for complex cases
Draeger sampling tubes and systems allow identification and measurement of different substances even under difficult conditions. Use Draeger Tubes to simultaneously and accurately measure very complex mixtures of substances or cases in which several similar substances are present. Dräger Tube Perchloroethylene 2/a has the Draeger SKU 8101501.
Dräger accuro
Fast measurement with one hand: The Dräger-Tube pump accuro allows you to use the established Dräger-Tubes to take measurements under extreme conditions. The easy to use Draeger Tube 8101501 Perchloroethylene 2/a has already undergone calibration. It forms a perfect unit with the Dräger-Tube pump including the Draeger Safety Perchloroethylene Tube.
- Dräger Accuro Pump info here >>
- Draeger Hard-Side Accuro Pump Kit info here >>
Perchloroethylene Draeger Tubes & CMS Handbook | App Store – Draeger Tubes App | Play Store – Draeger Tubes App |
What are Draeger Tubes?
Dräger-Tubes are glass vials that contain a chemical reagent that reacts to a specific chemical or family of chemicals. Draeger detector tubes are a cost-effective, fast, and reliable way to measure toxic gases anywhere.
Draeger Tube Perchloroethylene 2/a Gas Detection Tube 8101501
For more information about this product, contact us today +1(805) 617-0987.
Additional information
Weight | 1 lbs |
---|---|
Dimensions | 7 × 2 × 2 in |