Draeger Tube Acetic Acid 5/a 6722101
$94.50
The Draeger Tube Acetic Acid 5/a 6722101 has a measurement range of 5 – 80 ppm. During measurements color changes on the tube indicate the concentration level of the hazardous substance.
- Box of 10 tubes
- 2 year shelf life
Description
The Draeger Tube Acetic Acid 5/a 6722101 is high precision. The Dräger sampling tubes and systems provide very reliable results, even with complex compounds and mixtures of substances.
Dräger-Tubes for Measuring Air Samples
Application Range
- Standard Measuring Range: 5 to 80 ppm
- Number of Strokes n: 3
- Time for Measurement: approx. 30 s
- Standard Deviation: ± 10 to 15 %
- Color Change: blue violet → yellow
Ambient Operating Conditions
- Temperature: 10 to 40 °C
- Absolute Humidity: < 30 mg H2O / L
Reaction Principle
- CH3COOH + pH Indicator → yellow reaction product
Cross Sensitivity
It is impossible to measure acetic acid in the presence of other acids.
This tube indicates organic acids by the same color change, but with different sensitivities.
The tube indicates mineral acids (e.g. hydrochloric acid) by red discolorations and different sensitivities.
Dräger-Tubes for Measurements in Liquid Samples
Application Range
Determination of acetic acid in water/waste water
- Dräger-Tube: Acetic Acid 5/a
- Measuring range: 0.5 to 20 mg/L
- Number of Strokes (n): 10
- Typical Stroke Time: 10 to 30 s
- Measurement Time: approx. 200 s
- Sample Volume: 200 mL
- Color Change: blue violet → yellow
- Temperature Range: 10 to 30 °C
- pH-Measurement: necessary
Information of Measurement
Using sulfuric acid, the pH-value has to be adjusted to the value of 1.3
System Parameters
Measurement Range [mg/L] | Standard Deviation [%] | Temperature [°C] | Parameters B C |
0.5 to 20 | 25 | 10 to 30 |
0.339 1.368 |
Evaluation of Measurement
Calculate acetic acid concentration:
Y[g/L] = A • B • (X[ppm] + C)
Cross Sensitivity
Formic acid is indicated with lower and propionic acid with higher sensitivity. Draeger Tube Acetic Acid 0.5 to 20 g/L 6722101
Application Range
Determination of acetic acid in water/waste water
- Dräger-Tube: Acetic Acid 5/a
- Measuring range: 1 to 20 mg/L
- Number of Strokes (n): 10
- Typical Stroke Time: 10 to 30 s
- Measurement Time: approx. 200 s
- Sample Volume: 200 mL
- Color Change: blue violet → yellow
- Temperature Range: 5 to 25 °C
- pH-Measurement: necessary
Information of Measurement
Using sulfuric acid, the pH-value has to be adjusted to the value of 1.3
System Parameters
Measurement Range [mg/L] | Standard Deviation [%] | Temperature [°C] | Parameters B C |
1 to 20 | 25 | 5 to 25 |
0.487 1.607 |
Evaluation of Measurement
Calculate formic acid concentration:
Y[g/L] = A • B • (X[ppm] + C)
Cross Sensitivity
Acetic acid and propionic acid are indicated with higher sensitivity. Draeger Tube Formic Acid 1 to 20 g/L 6722101
Application Range
Determination of sum parameter acetic acid, formic acid and propionic acid in water/waste water
- Dräger-Tube: Acetic Acid 5/a
- Measuring range: 0.3 to 15 mg/L
- Number of Strokes (n): 10
- Typical Stroke Time: 10 to 30 s
- Measurement Time: approx. 200 s
- Sample Volume: 200 mL
- Color Change: blue violet → yellow
- Temperature Range: 10 to 25 °C
- pH-Measurement: necessary
Information of Measurement
Using sulfuric acid, the pH-value has to be adjusted to the value of 1.3. Draeger Tube Organic Acids 0.5 to 15 g/L 6722101
System Parameters
Measurement Range [mg/L] | Standard Deviation [%] | Temperature [°C] | Parameters B C |
0.5 to 15 | 25 | 10 to 25 |
0.241 1.157 |
Evaluation of Measurement
Calculate acid concentration:
Y[g/L] = A • B • (X[ppm] + C)
Application Range
Determination of propionic acid in water/waste water
- Dräger-Tube: Acetic Acid 5/a
- Measuring range: 0.3 to 10 mg/L
- Number of Strokes (n): 10
- Typical Stroke Time: 10 to 30 s
- Measurement Time: approx. 200 s
- Sample Volume: 200 mL
- Color Change: blue violet → yellow
- Temperature Range: 10 to 30 °C
- pH-Measurement: necessary
Information of Measurement
Using sulfuric acid, the pH-value has to be adjusted to the value of 1.3
System Parameters
Measurement Range [mg/L] | Standard Deviation [%] | Temperature [°C] | Parameters B C |
0.3 to 10 | 25 | 10 to 30 |
0.153 0.687 |
Evaluation of Measurement
Calculate propionic acid concentration:
Y[g/L] = A • B • (X[ppm] + C)
Cross Sensitivity
Acetic acid and formic acid are indicated with lower sensitivity. Draeger Tube Propionic Acid 0.3 to 10 g/L 6722101
Draeger Acetic Acid 5/a Tubes – 6722101
- Simple to use
- Requires no calibration
- Inexpensive and portable
- Wide variety of gases and vapors
- 70 years of proven technology
- Most Trusted – Draeger-Tubes have been leading the way in gas detection for over 60 years.
- Widest Range – Over 200 different Draeger-Tubes are available for measuring over 500 gases and vapors.
- Hassle-Free Operation – No calibration, no lubrication, no special maintenance. Always ready to go.
Accurate and reliable results
Draeger-Tubes with the Accuro Pump deliver the most accurate results. 60+ years of manufacturing colorimetric tubes. The bellows pump design delivers consistent volume. Individual calibration of each batch of Draeger gas detection tubes Acetic Acid included, accomplishes quality assurance. Every batch undergoes testing at regular intervals, throughout the two year shelf life, to guarantee accuracy over the entire life of the tubes.
Draeger Acetic Acid Detector Tubes
Collect the hazardous substances in the air by using a suitable medium such as activated carbon, silica gel, or molecular sieves. Then analyze the sample in a laboratory with the aid of instrumental analysis.
Suitable for complex cases
Draeger sampling tubes and systems allow identification and measurement of different substances even under difficult conditions. Use Draeger Tubes to simultaneously and accurately measure very complex mixtures of substances or cases in which several similar substances are present. Dräger Tube Acetic Acid 5/a has the Draeger SKU 6722101.
Dräger accuro
Fast measurement with one hand: The Dräger-Tube pump accuro allows you to use the established Dräger-Tubes to take measurements under extreme conditions. The easy to use Draeger Tube 6722101 Acetic Acid 5/a has already undergone calibration. It forms a perfect unit with the Dräger-Tube pump including the Draeger Safety Acetic Acid Tube.
- Dräger Accuro Pump info here >>
- Draeger Hard-Side Accuro Pump Kit info here >>
Acetic Acid Draeger Tubes & CMS Handbook | App Store – Draeger Tubes App | Play Store – Draeger Tubes App |
What are Draeger Tubes?
Dräger-Tubes are glass vials that contain a chemical reagent that reacts to a specific chemical or family of chemicals. Draeger detector tubes are a cost-effective, fast, and reliable way to measure toxic gases anywhere.
Draeger Tube Acetic Acid 5/a Gas Detection Tube 6722101
For more information about this product, contact us today +1(805) 617-0987.
Additional information
Weight | 1 lbs |
---|---|